________________________________________________________
LIFELINE AED Specifications*
D E F I B R I L L A T O R
S E L F  T E S T S
Type
Semi-automatic external
defibrillator
Model
DDU-100A
Waveform
Biphasic Truncated
exponential
Energy
150-Joules (Nominal
into 50 ohm load)
Charge Time(typical)
Less than 6
seconds(DBP-2800)
Less than 9
seconds(DBP-1400)
Shock-to-shock Cycle
Time
 Less than 20
seconds(typical includes
analysis and charging
time)
Voice Prompts
Extensivevoice prompts
guide user through
operation
Controls
-Lighted ON/Off switch
-Lighted shock button
Indicators
-"check pads"
-"do not touch patient"
-"analyzing"
-AED status LED
Automatic
Automatic daily,
weekly and monthly
circuitry tests
Battery Insertion
System integrity test
on battery insertion
User Initiated
Unit and battery pack
system test initiated by user

Status Indication
Visual and Audible
indication of the unit status
D E F I B R I L L A T I O N   P A D S
Model DDP-100
Type
Single-use,
disposable,
self-adhesive
electrodes with cable
and connector
Surface Area
103 cm2(nominal)
Pad Placement
Anterior-Anterior
Cable Length
48 in (122cm)
PATIENT ANALYSIS SYSTEM
Patient Analysis
Automatically evaluates
patient impedance for
proper pad contact.  
Monitors signal quality
and analyzes patient
ECG for
shockable/non-shockable
rythms
Sensitivity/Specificity
Meets AAMI-DF-39
specifications and AHA
recommendations
EVENT  DOCUMENTATION
Internal Event Record
Critical ECG segments
and rescue event
parameters are recorded
and can be downloaded
to a removable data card
PC Based Review
ECG with event tag
display, and audio
playback when available
Removable Storage
Up to 12 hours of
ECG and event data
storage (no audio) or
up to 1:40 of audio,
ECG and event
storage on a
removable data card.  
Actual length of
storage is dependent
on card capacity
_________________________________
B A T T E R Y   P A C K
E N V I R O N M E N T A L
Model DBP-1400
Power
15V, 1400 mAh
Capacity(new @25ºc)
125 shocks or 8 hours
continous operation
Standby-Life(typical)
5 years
Model DBP-2800
Power
15V, 2800mAh
Capacity(new @25ºC)
300 shocks or 16 hour
continuous operation
Standby-Life(typical)
7 Years
Temperature
Operating/Standby:
0-50ºC
(32-122ºF)
Humidity
Operating/Standby:
5%-95%
(non-condensing)
Altitude
-500 to 15,000 ft. per
MIL-STD-810F 500.4
Procedure II
Vibration

Ground
(MIL-STD-810F 514.5
Category 20)

Helicopter
(RTCA/DO-160D, Sec
8, Cat H, Zone 2,
Curve G)
Jet Aircraft
(RTCA/DO-160D, Sec 8,
Cat H, Zone 2, Curves B&R)
Shock/Drop Abuse
Tolerance
MIL-STD 810F 516.5
Procedure IV(1meter, any
edge, corner, or surface, in
stand-by mode)
Sealing/Water Resistance
IEC60529 class IP54; Splash
Proof, Dust Protected
(Battery Pack Installed.)
ESD
EN61000-4-2: 1998, (open
air up to 8kV or direct
contact up to 6kV)
EMC(emission)
EN60601-1-2 limits(1993),
method EN55011: 1998
Group 1 Level B
EMC(Immunity)
EN60601-1-1
limits(1993),method
EN61000-4-3 1998 Level 3
(10V/m)
TYPE                                   Low Battery Indicatiors
Lithium/Maganese                  -Visible
Dioxide                                   -Audible
Disposable, recyclable       
non-rechargable
P H Y S I C A L
Size
8.5 x 11.8 x 2.7
inches
(22 x 30 x 7 cm)
Weight
(Approximate)with
DBP-1400 4.2 lbs.
with DBP-2800 4.4 lbs.
*Specifications subject to change without notice


HOME   DEFIBRILLATORS   AED ACCESSORIES   FIRST-AID KITS   FLASHLIGHTS   CONTACT US   STORE